Product preview
Product preview

High temperature furnace LHTCT 03/16/C550 with controller C 550, SiC rod heated, max. 1550°C

Part #: 9.764 587Mfr #: LC034H6ON5HS Code #:85141980
Estimated shipping time:Contact us via live chat or via contact us page

RM 62,964.00

See the available (4) variant products
Specifications
  • Type/ControllerLHTCT 03/16/C550*
  • Capacity litres3.0
  • Temp. max. °C1600
  • External dimensions (W x D x H) mm415 x 545 x 490
  • Internal dimensions (W x D x H) mm120 x 210 x 120
  • Rating kW9.0
Downloads
  • Nabertherm HTCT 01_14 - HTCT 01_16 Instruction ManualDownload
  • Nabertherm B400-B410_C440-C45_P470_P480 Controller Instruction ManualDownload
Product overview
These powerful laboratory muffle furnaces are available for temperatures up to 1400 °C, 1500 °C, 1550 °C or 1600 °C. The durability of the SiC rods in periodic use, in combination with their high heating speed, make these furnaces the all-rounders in the laboratory. Heating times of 40 minutes to 1400 °C can be achieved, depending on the furnace
model and the conditions of use.
  • Tmax 1400 °C, 1500 °C, 1550 °C or 1600 °C

  • Working Temperature 1500 °C (for models LHTC ../16), increased wear and tear of heating elements must be expected in case of working at higher temperatures

  • Insulation only made of materials that are classified as not carcinogenic acc. to TRGS 905, class 1 or 2

  • Housing made of sheets of textured stainless steel

  • Dual shell housing for low external temperatures and high stability

  • Furnace room lined with durable fiber material, not classified

  • Optional flap door (LHTC) which can be used as work platform or lift door (LHTCT) with hot surface facing away from the operator (HTCT 01/ only with lift door)

  • Switching system with solid-state-relays, power tuned to the SiC rods

  • Controller is the programmable C 550 with adjustable ramps and holding times

  • Easy replacement of heating rods

  • Adjustable air inlet opening, exhaust air opening in the roof

  • 400 V 3/N/PE 50/60 Hz connector as standard


Option:
  • Over-temperature limiter with adjustable cutout temperature for thermal protection class 2 in accordance with EN 60519-2 as temperature limiter to protect the furnace and load

  • Square saggar for charging of up to three layers

  • Lid for top saggar

  • Manual or automatic gas supply system

  • Process control and documentation via VCD software package for monitoring, documentation and control
Variant products

Products related to this item (4)

Just add items to your shopping cart and select “Instant Quote”